Medical policy: External Infusion Pumps for Insulin Delivery and Automated Insulin Delivery Systems
Policy number: MP 6.007
Clinical benefit
- Minimize safety risk or concern.
- Minimize harmful or ineffective interventions.
- Assure appropriate level of care.
- Assure appropriate duration of service for interventions.
- Assure that recommended medical prerequisites have been met.
- Assure appropriate site of treatment or service.
Effective date: 5/1/2026
Policy
External insulin infusion pumps for treatment of insulin dependent diabetes mellitus may be considered medically necessary in individuals who cannot be controlled by intermittent dosing. (Acceptable routes are SC and IV)
Automated insulin delivery systems (a.k.a. artificial pancreas device systems)
Use of a U.S. Food and Drug Administration (FDA) cleared or approved automated insulin delivery system (artificial pancreas device system) with a low glucose suspend feature may be considered medically necessary in individuals with type 1 diabetes who meet all of the following criteria:
Age 6 years and older AND
- Glycated hemoglobin level between 5.8% and 10.0%;
- Used insulin pump therapy for more than 6 months; and
- At least 2 documented nocturnal hypoglycemic events in a 2-week period.
Use of an FDA cleared or approved automated insulin delivery system (artificial pancreas device system) designated as a hybrid closed-loop insulin delivery system (with low glucose suspend and suspend before low features) may be considered medically necessary in individuals with type 1 diabetes who meet all of the following criteria:
Over age 6 years AND
- Glycated hemoglobin level between 5.8% and 10.0%;
- Used insulin pump therapy for more than 6 months; and
- At least 2 documented nocturnal hypoglycemic events in a 2-week period.
OR
Age 2 to 6 years AND
- Clinical diagnosis of type 1 diabetes for 3 months or more;
- Used insulin pump therapy for more than 3 months;
- Glycated hemoglobin level <10.0%;
- Minimum daily insulin requirement (Total Daily Dose) of greater than or equal to 8 units.
Use of an FDA cleared or approved automated insulin delivery system (artificial pancreas device system) designated as a closed-loop insulin delivery system may be considered medically necessary in individuals with type 1 diabetes who meet all of the following criteria:
Age 6 years and older AND
- Clinical diagnosis of type 1 diabetes for 12 months or more;
- Using insulin for at least 12 months;
- Diabetes managed using the same regimen (either pump or multiple daily injections, with or without continuous glucose monitoring) for 3 months or longer.
Use of an FDA cleared or approved automated insulin delivery system (artificial pancreas device system) designated as a hybrid closed-loop insulin delivery system (with low glucose suspend and suspend before low features) may be considered medically necessary in individuals with type 2 diabetes who meet all the following criteria: (See policy guidelines)
Age 18 years and older AND
- Diagnosed with type 2 diabetes for at least 12 months;
- On multiple daily injections (insulin administration ≥ 3x/day) or use of insulin infusion pump for at least 3 months;
- Glycated hemoglobin level ≥7% or experience significant hypoglycemia.
Use of an automated insulin delivery system (artificial pancreas device system) is considered investigational for individuals who do not meet the above criteria as there is insufficient evidence to support a general conclusion concerning the health outcomes or benefits associated with this device for any other indications.
Use of an automated insulin delivery system (artificial pancreas device system) not cleared or approved by the FDA is investigational as there is insufficient evidence to support a general conclusion concerning the health outcomes or benefits associated with this device for any other indications.
Note: A programmable disposable external insulin infusion pump (e.g., OmniPod®) is an acceptable alternative to a standard insulin infusion pump.
External infusion pump/automated insulin delivery system replacement
Requests for replacement of an insulin pump that is out of warranty must include one of the following:
- Clear and conclusive documentation from either the treating physician’s office notes or the device supplier’s customer service notes, that the pump is non-operational; or
- Documentation that the individual has reverted to use of multiple daily injections of insulin or a loaner pump because the pump is non-operational.
Replacement of insulin pumps for reasons other than those stated above are considered investigational as there is insufficient evidence to support a general conclusion concerning the health outcomes or benefits associated with this device for any other indications.
Policy guidelines
Individuals with diabetes mellitus should understand the technology, be motivated to use the device correctly and consistently, adhere to a comprehensive treatment plan supervised by a qualified provider, and be capable of recognizing alerts and alarms from the device.
The U.S. Food and Drug Administration (FDA) has approved two hybrid closed-loop insulin delivery systems for type 2 diabetes management: the Omnipod® 5 AID system (Insulet Corporation) and t:slim X2 insulin pump equipped with Control-IQ+ technology (Tandem Diabetes Care).
- The Omnipod 5 AID system was approved following a pivotal clinical trial by Pasquel et al in 2025 (SECURE-T2D; NCT05815342). Trial participants aged 18 to 75 years, who had been on a stable insulin regimen for at least three months prior to screening (as per the above policy criteria), were selected. Additionally, they could be on other antihyperglycemic and weight loss drugs, provided there were no dose changes for at least 4 weeks before the trial commenced. However, those who experienced more than one severe hypoglycemic event or episode of diabetic ketoacidosis or hyperosmolar hyperglycemic syndrome within 6 months before screening were excluded.
- The t:slim X2 insulin pump equipped with Control-IQ+ technology was approved following a pivotal randomized controlled trial by Kudva et al in 2025 (2IQP; NCT0585832). Trial participants were aged 19 to 87 years and had type 2 diabetes for at least 6 months, according to clinical history and available laboratory data. All participants were receiving multiple daily injections of insulin with at least one injection containing rapid-acting insulin per day or were using an insulin pump for at least 3 months before enrollment. Mixed insulin use with a rapid-acting component was allowed. Concurrent treatment with noninsulin glucose-lowering medications or weight-reduction medications was permitted, provided the dose had been stable for the previous 3 months; during the trial, these medications were continued in both treatment groups.
Cross-references
- MP 6.004 Continuous Glucose Monitoring
Product variations
This policy is only applicable to certain programs and products administered by Capital Blue Cross and subject to benefit variations. Please see additional information below.
FEP PPO - Refer to the FEP Medical Policy Manual. The FEP medical policy manual can be found at FEP Medical Policy Manual.
Description/background
External infusion pump (EIP) is a portable battery-operated device intended to provide continuous ambulatory drug infusion therapy over an extended time period. The EIP is also known as an external pump, an ambulatory pump, or a mini-infuser. Proposed drug delivery routes using the EIP include the intravenous, intra-arterial, subcutaneous, intraperitoneal, epidural, intrathecal, and intraventricular routes. A heparinized saline solution may be used during an interruption of drug therapy to maintain catheter patency. A catheter from the pump is attached to the desired access route for drug delivery. The drug reservoir refilling is non-invasive.
There are over six hundred different models of pumps, most of which have received clearance for marketing by the Food and Drug Administration (FDA) through a pre-notification application (510(k)).
Automated insulin delivery systems
Automated insulin delivery systems (AID), also known as artificial pancreas device systems, have a continuous glucose monitor linked to an insulin pump with the capability to automatically stop, reduce, or increase insulin infusion based on specified thresholds of measured interstitial glucose.
AID components are designed to communicate with each other to automate the process of maintaining blood glucose concentrations at or near a specified range or target and to minimize the incidence and severity of hypoglycemic and hyperglycemic events. A control algorithm is embedded in software in an external processor or controller that receives information from the continuous glucose monitor and performs a series of mathematical calculations. Based on these calculations, the controller sends dosing instructions to the infusion pump.
Different AID types are currently available for clinical use. Sensor augmented pump therapy (SAPT) with low glucose suspend (LGS) or suspend on low may reduce the likelihood or severity of a hypoglycemic event by suspending insulin delivery temporarily when the sensor value reaches (reactive) a predetermined lower threshold of measured interstitial glucose. Low glucose suspension (LGS) automatically suspends basal insulin delivery for up to two hours in response to sensor-detected hypoglycemia.
A sensor augmented pump therapy with predictive low glucose management (PLGM) (suspend before low) suspends basal insulin infusion with the prediction of hypoglycemia. Basal insulin infusion is suspended when sensor glucose is at or within 70 mg/dL above the patient’s low limit and is predicted to be 20 mg/dL above this low limit in 30 minutes. In the absence of a patient response, the insulin infusion resumes after a maximum suspend period of two hours. In certain circumstances, auto-resumption parameters may be used.
When a sensor value is above or predicted to remain above this threshold, the infusion pump will not take any action based on continuous glucose monitor readings. Patients using this system still need to monitor their blood glucose concentration, set appropriate basal rates for their insulin pump, and give pre-meal bolus insulin to control their glucose levels.
A hybrid closed-loop system automatically increases or decreases basal insulin delivery in response to sensor glucose values. The user still needs to dose prandial insulin manually. Next-generation systems not only adjust basal insulin delivery but also have the capability to deliver automatic correction boluses. However, they still require the person with diabetes to dose prandial insulin.
A closed-loop system has automated insulin delivery and continuous glucose sensing without patient intervention. These systems utilize a control algorithm that autonomously and continually increases and decreases the subcutaneous insulin delivery based on real-time sensor glucose levels.
Table 1 summarizes the FDA cleared or approved automated insulin delivery systems.
Table 1. U.S. Food and Drug Administration-Approved Automated Insulin Delivery Systems (Artificial Pancreas Device Systems)
Device |
Age indication |
Manufacturer |
Date approved |
PMA no. / device code |
|
MiniMed 530G system1 (open-loop, LGS) |
≥16 years |
Medtronic |
July 2013 |
P120010 / OZO |
|
MiniMed 630G system with SmartGuard™2 (open-loop, LGS) |
≥16 years ≥14 years |
Medtronic |
Aug 2016 Jun 2017 |
P150001 / OZO P150001 / S008 |
|
MiniMed 670G system3 (HCL, LGS or PLGM) |
≥14 years ≥7 to 13 years |
Medtronic |
Sep 2016 Jul 2018 |
P160017 / OZP P160017 / S031 |
|
MiniMed 770G system4 (HCL) |
≥2 years |
Medtronic |
Aug 2020 |
P160017 / S076 |
|
MiniMed 780G system5 (HCL) |
≥7 years |
Medtronic |
May 2023 |
P160017 / S091 |
|
t:slim X2 insulin pump with Basal-IQ technology (LGS) |
≥6 years |
Tandem |
Jun 2018 |
P180008 / OZO, PQF |
|
t:slim X2 insulin pump with Control-IQ technology (HCL) |
≥6 years |
Tandem |
Dec 2019 |
DEN180058 / QFG |
|
Omnipod 5 (HCL) |
≥6 years |
Insulet |
Jan 2022 |
K203768 K203772 |
|
iLet bionic pancreas (CL) |
≥6 years |
Beta Bionics |
May 2023 |
K220916 K223846 |
CL: closed-loop; HCL: hybrid closed-loop; LGS: low glucose suspend; OZO: Artificial Pancreas Device System, threshold suspend; OZP: Automated Insulin Dosing Device System, Single Hormonal Control; PMA: premarket approval; PLGM: predictive low glucose management.
1MiniMed 530G System consists of the following devices that can be used in combination or individually: MiniMed 530G Insulin Pump, Enlite™ Sensor, Enlite™Serter, the MiniLink Real-Time System, the Bayer Contour NextLink glucose meter, CareLink® Professional Therapy Management Software for Diabetes, and CareLink® Personal Therapy Management Software for Diabetes (at time of approval).
2MiniMed 630G System with SmartGuard™ consists of the following devices: MiniMed 630G Insulin Pump, Enlite® Sensor, One-Press Serter, Guardian® Link Transmitter System, CareLink® USB, Bayer’s CONTOUR® NEXT LINK 2.4 Wireless Meter, and Bayer’s CONTOUR® NEXT Test Strips (at time of approval).
3MiniMed 670G System consists of the following devices: MiniMed 670G Pump, the Guardian Link (3) Transmitter, the Guardian Sensor (3), One-Press Serter, and the Contour NEXT Link 2.4 Glucose Meter (at time of approval).
4MiniMed 770G System consists of the following devices: MiniMed 770G Insulin Pump, the Guardian Link (3) Transmitter, the Guardian Sensor (3), One-Press Serter, the Accu-Chek Guide™ Link blood glucose meter, and the Accu-Chek Guide™ Test Strips.
5MiniMed 780G System consists of the following devices: MiniMed 780G Insulin Pump, the Guardian 4 Transmitter, the Guardian 4 Sensor (3), One-Press Serter, the Accu-Chek Guide™ Link blood glucose meter, and the Accu-Chek Guide™ Test Strips.
The MiniMed 530G system includes a threshold suspend or low glucose suspend feature. The threshold suspend tool temporarily suspends insulin delivery when the sensor glucose level is at or below a preset threshold within the 60- to 90-mg/dL range. When the glucose value reaches this threshold, an alarm sounds. If patients respond to the alarm, they may choose to continue or cancel the insulin suspend feature. If patients fail to respond, the pump automatically suspends action for 2 hours, and then insulin therapy resumes.
The MiniMed 630G system with SmartGuard™, which is similar to the 530G, includes updates to the system components including waterproofing. The threshold suspend feature can be programmed to temporarily suspend delivery of insulin for up to 2 hours when the sensor glucose value falls below a predefined threshold value. The MiniMed 630G system with SmartGuard™ is not intended to be used directly for making therapy adjustments, but rather to provide an indication of when a finger stick may be required. All therapy adjustments should be based on measurements obtained using a home glucose monitor and not on the values provided by the MiniMed 630G system. The device is not intended to be used directly for preventing or treating hypoglycemia but to suspend insulin delivery when the user is unable to respond to the SmartGuard™ Suspend on Low alarm to take measures to prevent or treat hypoglycemia themselves.
The MiniMed 670G system is a hybrid closed-loop insulin delivery system consisting of an insulin pump, a glucose meter, and a transmitter, linked by a proprietary algorithm and the SmartGuard Hybrid closed-loop. The system includes a low glucose suspend feature that suspends insulin delivery; this feature either suspends delivery on low-glucose levels or suspends delivery before low-glucose levels and has an optional alarm (manual mode). Additionally, the system allows semiautomatic basal insulin-level adjustment (decrease or increase to preset targets) (automatic mode). As a hybrid system, basal insulin levels are automatically adjusted, but the patient needs to administer premeal insulin boluses. The continuous glucose monitoring component of the MiniMed 670G system is not intended to be used directly for making manual insulin therapy adjustments; rather it is to provide an indication of when a glucose measurement should be taken. The MiniMed 670G system was originally approved for marketing in the United States on September 28, 2016 (P160017) and received approval for marketing with a pediatric indication (ages 7 to 13 years) on June 21, 2018 (P160017/S031).
The MiniMed 770G system is an iteration of the MiniMed 670G system. In July 2020, the device was approved for use in children ages 2 to 6 years. In addition to the clinical studies that established the safety and effectiveness of the MiniMed 670G system in users ages 7 years and older, the sponsor performed clinical studies of the 670G system in pediatric subjects ages 2 to 6 years. The FDA concluded that these studies establish a reasonable assurance of the safety and effectiveness of the MiniMed 770G system because the underlying therapy in the 670G system and the associated Guardian Sensor 3 are identical to that of the 770G system.
On June 21, 2018, the FDA approved the t:slim X2 insulin pump with Basal-IQ technology (PMA P180008) for individuals who are 6 years of age and older. The system consists of the t:slim X2 insulin pump paired with the Dexcom G5 mobile continuous glucose monitoring system, as well as the Basal-IQ technology. The t:slim X2 insulin pump is intended for the subcutaneous delivery of insulin, at set and variable rates, for the management of diabetes mellitus in persons requiring insulin. The t:slim X2 insulin pump can be used solely for continuous insulin delivery and as part of the system as the receiver for therapeutic continuous glucose monitoring. The t:slim X2 insulin pump running the Basal-IQ technology can be used to suspend insulin delivery based on continuous glucose monitoring sensor readings.
In December 2019, FDA approved the t:slim X2 insulin pump with Control-IQ technology through the De Novo process. The device uses the same pump hardware as the insulin pump component of the system approved in the t:slim X2 insulin pump with Basal-IQ technology (P180008) and P140015. A custom disposable cartridge is motor driven to deliver patient programmed basal rates and boluses through an infusion set into subcutaneous tissue.
In 2022, FDA approved the Omnipod 5 AID system for the subcutaneous delivery of insulin, at set and variable rates, for the management of diabetes mellitus in persons requiring insulin. The Omnipod 5 AID system is able to reliably and securely communicate with compatible, digitally connected devices, including automated insulin dosing software, to receive, execute, and confirm commands from these devices.
In May 2023, FDA approved the first closed-loop system through the 510(k) premarket clearance pathway.
Rationale
Summary of evidence: External infusion pumps for insulin delivery
For individuals who have type 1 or type 2 diabetes who receive insulin therapy via an external infusion pump, the American Diabetes Association Standards of Care in Diabetes 2024 states that insulin pumps have been available in the United States for over 40 years. These devices deliver rapid-acting insulin throughout the day to help manage glucose levels. Most studies that compare multiple daily injections with insulin pump therapy have been relatively small and of short duration. However, a systematic review and meta-analysis concluded that pump therapy has modest advantages for lowering A1C levels and for reducing severe hypoglycemia rates in children and adults. Real world data on insulin pump use in individuals with type 1 diabetes show benefits in A1C levels and hypoglycemia reduction as well as total daily insulin dose reduction.
Real-world reports have shown reduction of A1C levels and reduction of total daily insulin dose in individuals with type 2 diabetes initiating insulin pump therapy.
Due to societal guidance, the evidence is sufficient to determine that the technology results in an improvement in the net health outcome.
Summary of evidence: Automated insulin delivery systems
For individuals who have type 1 diabetes who receive an automated insulin delivery system with a low glucose suspend feature, the evidence includes 3 randomized controlled trials, conducted in some settings. Relevant outcomes include symptoms, changes in disease status, morbid events, resource utilization, and treatment-related morbidity. Primary eligibility criteria of the key RCT, the Automate to Simulate Pancreatic Insulin Response (ASPIRE) trial, were ages 16 to 70 years, type 1 diabetes, glycated hemoglobin levels between 5.8% and 10.0%, and at least 2 nocturnal hypoglycemic events (≤65 mg/dL) lasting more than 20 minutes during a 2-week run-in phase. Both trials required at least 6 months of insulin pump use. Both RCTs reported significantly less hypoglycemia in the treatment group than in the control group. In both trials, primary outcomes were favorable for the group using an automated insulin delivery system; however, findings from 1 trial were limited by nonstandard reporting of hypoglycemic episodes, and findings from the other trial were no longer statistically significant when the number of enrollees (children) were excluded from analysis. The RCT limited to adults showed an improvement in the primary outcome (area under the curve for nocturnal hypoglycemic events). The area under the curve is not used for assessment in clinical practice but the current technology does allow user and provider review of similar data with continuous glucose monitoring. Results from the ASPIRE study suggested that there were increased risks of hyperglycemia and potential diabetic ketoacidosis in subjects using the threshold suspend feature. This finding may be related to whether or not actions are taken by the user to assess glycemic status, the etiology of low glucose readings (activity, diet, medication), or to reduction in insulin infusion. Both retrospective and prospective observational studies have reported reductions in rates and severity of hypoglycemic episodes with automated insulin delivery systems. The evidence suggests that the magnitude of reduction of hypoglycemic events in the type 1 diabetes population is likely to be clinically significant. The evidence is sufficient to determine that the technology results in an improvement in the net health outcome.
For individuals who have type 1 diabetes who receive an automated insulin delivery device system with a hybrid closed-loop insulin delivery system, the evidence includes multicenter pivotal trials using devices cleared by the U.S. Food and Drug Administration, supplemental data and analysis for expanded indications, and more recent studies focused on children and adolescents. Three crossover randomized controlled trials using a similar first-generation device studied and approved outside the United States have been reported. Relevant outcomes are symptoms, change in disease status, morbid events, resource utilization, and treatment-related morbidity. Of these 3 crossover RCTs, 2 found significantly better outcomes (i.e., time spent in nocturnal hypoglycemia and time spent in preferred glycemic range) with the device than with standard care. The third study had mixed findings (significant difference in time spent in nocturnal hypoglycemia and no significant difference in time spent in preferred glycemic range). Additional evidence from device performance studies and clinical studies all demonstrate gains in time spent in various levels of hypoglycemia, improved time in range (70-180 mg/dL), rare diabetic ketoacidosis, and few device-related adverse events. The evidence suggests that the magnitude of reduction of hypoglycemic events in the type 1 diabetes population is likely to be clinically significant. The evidence is sufficient to determine that the technology results in an improvement in the net health outcome.
For individuals who have type 1 diabetes who receive an automated insulin delivery system with a closed-loop insulin delivery system, the evidence includes a 13-week multicenter randomized controlled trial of the iLet bionic pancreas system compared with usual care in 326 individuals ages 6 to 79 years with type 1 diabetes. Comparator group participants continued their pre-study subcutaneous insulin delivery (either multiple daily injections, an insulin pump without an automated insulin delivery system, or an insulin pump with predictive low glucose suspend feature, or an insulin pump as part of an artificial pancreas system) plus real-time continuous glucose monitoring. The glycated hemoglobin level decreased from 7.9% to 7.3% in the closed-loop insulin delivery system group and did not change (7.7%) in the control group (mean adjusted difference at 13 weeks, −0.5%; 95% CI −0.6 to −0.3; p<0.001). The rate of severe hypoglycemia was 17.7 events per 100 participant-years in the closed-loop insulin delivery system group and 10.8 events per 100 participant-years in the standard-care group (p = 0.39). No episodes of diabetic ketoacidosis occurred in either group. The trial results for the subgroups of adults (ages 18 and older) and youth (ages 6 to 17 years) have additionally been reported and were similar to the main results for the full cohort. The evidence is sufficient to determine that the technology results in an improvement in the net health outcome.
Additional information
2019 input
Clinical input supported that the outcome of hypoglycemia prevention provides a clinically meaningful improvement in net health outcome, and this use is consistent with generally accepted medical practice. Clinical input also supported that the use of hybrid closed-loop artificial pancreas device systems provides a clinically meaningful improvement in net health outcome and is consistent with generally accepted medical practice. Reduction in the experience of hypoglycemia and inappropriate awareness of hypoglycemia and glycemic excursions were identified as important acute clinical outcomes in children, adolescents, and adults and are related to the future risk of end-organ complications.
Definitions
N/A
Disclaimer
Capital Blue Cross’ medical policies are used to determine coverage for specific medical technologies, procedures, equipment, and services. These medical policies do not constitute medical advice and are subject to change as permitted by law or applicable clinical evidence from independent treatment guidelines. Treating providers are solely responsible for medical advice and treatment of members. These policies are not a guarantee of coverage or payment. Payment of claims is subject to a determination regarding the member’s benefit program and eligibility on the date of service, and a determination that the services are medically necessary and appropriate. Final processing of a claim is based upon the terms of contract that applies to the member’s benefit program, including benefit limitations and exclusions. If a provider or a member has a question concerning this medical policy, please contact Capital Blue Cross’ Provider Services or Member Services.
Coding information
Note: This list of codes may not be all-inclusive, and codes are subject to change at any time. The identification of a code in this section does not denote coverage as coverage is determined by the terms of member benefit information. In addition, not all covered services are eligible for separate reimbursement.
Covered when medically necessary
Procedure codes |
||||
|
A4224 |
A4225 |
A4226 |
A4230 |
A4231 |
|
A4232 |
A9274 |
E0784 |
E0787 |
S1034 |
|
S1035 |
S1036 |
S1037 |
|
|
*Specific ICD-10-CM codes do not apply; must meet policy criteria above.
References
- Eugster EA, Francis G. Lawson-Wilkins Drug and Therapeutics Committee. Position statement: continuous subcutaneous insulin infusion in very young children with type I diabetes. Pediatrics. 2006;118:1244-1249.
- Levitsky L, Misra M. Overview of the management of type 1 diabetes mellitus in children and adolescents. In: UpToDate Online Journal [serial online]. Waltham, MA: UpToDate; updated November 15, 2023. Literature review current through November 2024.
- Weinstock R. Management of blood glucose in adults with type 1 diabetes mellitus. In: UpToDate Online Journal [serial online]. Waltham, MA: UpToDate; updated January 2, 2024. Literature review current through November 2024.
- Wexler D. Insulin therapy in type 2 diabetes mellitus. In: UpToDate Online Journal [serial online]. Waltham, MA: UpToDate; updated October 22, 2024. Literature review current through November 2024.
- Durnwald C. Gestational diabetes mellitus: glucose management and maternal prognosis. In: UpToDate Online Journal [serial online]. Waltham, MA: UpToDate; updated July 5, 2024. Literature review current through November 2024.
- Weinstock R. Continuous subcutaneous insulin infusion (insulin pump). In: UpToDate Online Journal [serial online]. Waltham, MA: UpToDate; updated June 7, 2024. Literature review current through November 2024.
- American Diabetes Association. Glycemic targets. Diabetes Care. Jan 2017;40(Suppl 1):S48-S56. PMID 27979893
- American Diabetes Association. Glycemic targets: standards of medical care in diabetes-2019. Diabetes Care. Jan 2019;42(Suppl 1):S61-S70. PMID 30559232
- Abraham MB, Jones TW, Naranjo D, et al. ISPAD Clinical Practice Consensus Guidelines 2018: assessment and management of hypoglycemia in children and adolescents with diabetes. Pediatr Diabetes. Oct 2018;19 Suppl 27:178-192. PMID 29863538
- Agiostratidou G, Anhalt H, Ball D, et al. Standardizing clinically meaningful outcome measures beyond HbA1c for type 1 diabetes: a consensus report of the American Association of Clinical Endocrinologists, the American Association of Diabetes Educators, the American Diabetes Association, the Endocrine Society, JDRF International, The Leona M. and Harry B. Helmsley Charitable Trust, the Pediatric Endocrine Society, and the T1D Exchange. Diabetes Care. Dec 2017;40(12):1622-1630. PMID 29162582
- Food and Drug Administration (FDA). Guidance for industry and Food and Drug Administration staff: content of investigational device exemption (IDE) and premarket approval (PMA) applications for artificial pancreas device systems [draft]. 2012.
- Food and Drug Administration. MiniMed 770G System. Summary of safety and effectiveness data. 2020.
- Food and Drug Administration. 2023. MiniMed 780G System - P160017/S091.
- Forlenza GP, Li Z, Buckingham BA, et al. Predictive low-glucose suspend reduces hypoglycemia in adults, adolescents, and children with type 1 diabetes in an at-home randomized crossover study: results of the PROLOG trial. Diabetes Care. Oct 2018;41(10):2155-2161. PMID 30089663
- Food and Drug Administration. 2023. FDA clears new insulin pump and algorithm-based software to support enhanced automatic insulin delivery.
- Food and Drug Administration (FDA). Premarket approval (PMA): MiniMed 530G system. 2013.
- Food and Drug Administration (FDA). Premarket approval (PMA): MiniMed 630G system with SmartGuard. 2016.
- Food and Drug Administration (FDA). Premarket approval (PMA): MiniMed 670G system. 2016.
- Food and Drug Administration (FDA). t:slim X2 insulin pump with Basal-IQ technology. Premarket approval. 2018.
- Faulds ER, Zappe J, Dungan KM. Real-world implications of hybrid close-loop (HCL) insulin delivery system. Endocr Pract. May 2019;25(5):477-484. PMID 30865545
- Bergenstal RM, Garg S, Weinzimer SA, et al. Safety of a hybrid closed-loop insulin delivery system in patients with type 1 diabetes. JAMA. Oct 04 2016;316(13):1407-1408. PMID 27629148
- Ly TT, Nicholas JA, Retterath A, et al. Effect of sensor-augmented insulin pump therapy and automated insulin suspension vs standard insulin pump therapy on hypoglycemia in patients with type 1 diabetes: a randomized clinical trial. JAMA. Sep 25 2013;310(12):1240-1247. PMID 24065010
- Forlenza GP, Eklundhammar L, Breton M, et al. Successful at-home use of the Tandem Control-IQ artificial pancreas system in young children during a randomized controlled trial. Diabetes Technol Ther. Apr 2019;21(4):159-169. PMID 30888835
- American Diabetes Association. Diabetes technology: standards of medical care in diabetes-2021. Diabetes Care. Jan 2021;44(Suppl 1):S85-S99. PMID 33298418
- Garg SK, Weinzimer SA, Tamborlane WV, et al. Glucose outcomes with the in-home use of a hybrid closed-loop insulin delivery system in adolescents and adults with type 1 diabetes. Diabetes Technol Ther. Mar 2017;19(3):155-163. PMID 28134564
- Beato-Víboras PI, Gallego-Gamero F, Lázaro-Martín L, et al. Prospective analysis of the impact of commercialized hybrid closed-loop system on glycemic control, glycemic variability, and patient-related outcomes in children and adults: a focus on superiority over predictive low-glucose suspend technology. Diabetes Technol Ther. Mar 2020;22(12):912-919. PMID 31855446
- Garg S, Brazg RL, Bailey TS, et al. Reduction in duration of hypoglycemia by automatic suspension of insulin delivery: the ASPIRE study. Diabetes Technol Ther. Mar 2012;14(3):205-209. PMID 22316089
- Forlenza GP, Deshpande S, Ly TT, et al. Application of zone model predictive control artificial pancreas during extended use of infusion set and sensor: a randomized crossover-controlled home-use trial. Diabetes Care. Aug 2017;40(8):1096-1102. PMID 28584076
- Pinsker JE, Dassau E, Deshpande S, et al. Outpatient randomized crossover comparison of zone model predictive control automated insulin delivery with weekly data-driven adaptation versus sensor-augmented pump: results from the international diabetes closed-loop trial 4. Diabetes Technol Ther. Sep 2022;24(9):635-642. PMID 35549708
- Tauschmann M, Thabit H, Bally L, et al. Closed-loop insulin delivery in suboptimally controlled type 1 diabetes: a multicentre, 12-week randomised trial. Lancet. Oct 13 2018;392(10155):1321-1329. PMID 30292578
- Abraham MB, Nicholas JA, Smith GJ, et al. Reduction in hypoglycemia with the predictive low-glucose management system: a long-term randomized controlled trial in adolescents with type 1 diabetes. Diabetes Care. Feb 2018;41(2):303-310. PMID 29191484
- Wood MA, Shulman DI, Forlenza GP, et al. In-clinic evaluation of the MiniMed 670G system "suspend before low" feature in children with type 1 diabetes. Diabetes Technol Ther. Nov 2018;20(11):731-737. PMID 30299976
- Breton MD, Kanapka LG, Beck RW, et al. A randomized trial of closed-loop control in children with type 1 diabetes. N Engl J Med. Aug 27 2020;383(9):836-845. PMID 32846062
- Kanapka LG, Wadwa RP, Breton MD, et al. Extended use of the Control-IQ closed-loop control system in children with type 1 diabetes. Diabetes Care. Feb 2021;44(2):473-478. PMID 33355258
- Cobry EC, Kanapka LG, Cengiz E, et al. Health-related quality of life and treatment satisfaction in parents and children with type 1 diabetes using closed-loop control. Diabetes Technol Ther. Jun 2021;23(6):401-409. PMID 33404325
- Forlenza GP, Pinhas-Hamiel O, Liljenquist DR, et al. Safety evaluation of the MiniMed 670G system in children 7–13 years of age with type 1 diabetes. Diabetes Technol Ther. Jan 2019;21(1):11-19. PMID 30585770
- Messer LH, Forlenza GP, Sherr JL, et al. Optimizing hybrid closed-loop therapy in adolescents and emerging adults using the MiniMed 670G system. Diabetes Care. Apr 2018;41(4):789-796. PMID 29444895
- Brown SA, Forlenza GP, Bode BW, et al. Multicenter trial of a tubeless, on-body automated insulin delivery system with customizable glycemic targets in pediatric and adult participants with type 1 diabetes. Diabetes Care. Jul 2021;44(7):1630-1634. PMID 34099518
- Russell SJ, Beck RW, Damiano ER, et al. Multicenter, randomized trial of a bionic pancreas in type 1 diabetes. N Engl J Med. Sep 29 2022;387(13):1161-1172. PMID 36170500
- Kruger D, Kass A, Lonier J, et al. A multicenter randomized trial evaluating the insulin-only configuration of the bionic pancreas in adults with type 1 diabetes. Diabetes Technol Ther. Oct 2022;24(10):697-711. PMID 36173236
- Messier L, Buckingham BA, Cogen FR, et al. Positive impact of the bionic pancreas on diabetes control in youth 6–17 years old with type 1 diabetes: a multicenter randomized trial. Diabetes Technol Ther. Oct 2022;24(10):712-725. PMID 36173237
- Lynch J, Kanapka LG, Russell SJ, et al. The insulin-only bionic pancreas pivotal trial extension study: a multi-center single-arm evaluation of the insulin-only configuration of the bionic pancreas in adults and youth with type 1 diabetes. Diabetes Technol Ther. Oct 2022;24(10):726-736. PMID 36173238
- Gruberger G, Sherr J, Allemann D, et al. American Association of Clinical Endocrinology clinical practice guideline: the use of advanced technology in the management of persons with diabetes mellitus. Endocr Pract. Jun 2021;27(6):505-537. PMID 34116789
- ElSayed NA, Aleppo G, Bannuru RR, et al. Diabetes technology: standards of care in diabetes—2024. Diabetes Care. Jan 01 2024;47(Suppl 1):S126-S144. PMID 38708755
- Brazeau AS, Kochhar KP, Raghinaru D, et al. Six-month randomized, multicenter trial of closed-loop control for type 1 diabetes. N Engl J Med. Oct 31 2019;381(18):1707-1717. PMID 31618560
- American Diabetes Association Professional Practice Committee. Diabetes technology: standards of care in diabetes—2025. Diabetes Care. Jan 01 2025;48(Suppl 1):S146-S166.
- Pasquel FJ, Davis GM, Huffman DM, et al. Automated insulin delivery in adults with type 2 diabetes: a nonrandomized clinical trial. JAMA Netw Open. Feb 03 2025;8(2):e2459348. PMID 39951268
- Kudva YC, Raghinaru D, Lum JW, et al. A randomized trial of automated insulin delivery in type 2 diabetes. N Engl J Med. Mar 19 2025. PMID 40105270
Policy history |
|
|
MP 6.007 |
03/05/2019 Consensus review. No change to policy statements. |
|
01/01/2020 Administrative update. New 2020 codes added to policy, A4225, A4226 and E0787. |
|
|
03/27/2020 Consensus review. No change to policy statement or coding. References updated. |
|
|
06/17/2021 Consensus review. Added guideline for hypoglycemia (“usually documented blood glucose levels less than 70 mg/dL and/or when an individual becomes symptomatic”) within policy statement. References updated. |
|
|
02/25/2022 Consensus review. No change to policy statements. References updated. HCPCS definitions removed. |
|
|
06/08/2022 Minor review. Artificial pancreas devices have been placed into this policy. Coding table updated. |
|
|
11/29/2022 Administrative update. Added new codes A4239 and E2103. |
|
|
12/14/2023 Minor review. Multiple daily injections are 3 or more per day. Deleted criteria re: self-adjustment for 6 months, which now allows for pumps at diagnosis. Deleted criteria re: T2D and needing to be on 2 oral drugs concomitantly (criteria is for any type of diabetes mellitus). Artificial pancreas system renamed automated insulin delivery system; changed title of policy. Modified criteria for automated insulin delivery system and also gave an allowance for those with type 2 diabetes. |
|
|
12/27/2024 Minor review. Automated insulin delivery system section criteria updated; allowance is only for those with type 1 diabetes. Updated references. No changes to coding. |
|
|
06/23/2025 Administrative update. Removed benefit variations section and updated disclaimer. |
|
|
11/20/2025 Minor review. Updated criteria for external infusion pumps. Automated insulin delivery systems now medically necessary for type 2 diabetes when medical necessity criteria is met. References updated. |
|